Interpretation of GB/T 18659-2002 for Electromagnetic Flowmeters

Tech Articles 2026-05-03By SIENCUN
GB/T 18659-2002 'Measurement of conductive liquid flow in closed conduits – Methods for evaluating the performance of electromagnetic flowmeters' is a key national standard. This article interprets its scope, referenced standards, definitions, performance indicators, test methods, and inspection rules, helping engineers and procurement professionals understand requirements and improve selection and application.

Introduction

GB/T 18659-2002, identical to ISO 9104:1991, specifies methods for evaluating the performance of electromagnetic flowmeters (EMFs) for conductive liquid flow in closed pipes. It provides unified evaluation criteria for manufacturers, users, and third-party testing bodies, crucial for measurement accuracy and reliability.

Scope

The standard applies to EMFs with integral or remote sensors and converters. Liquid conductivity is typically ≥5 μS/cm; lower conductivity requires technical agreements. It does not cover open-channel or insertion flowmeters.

Referenced Standards

  • GB/T 17614.1-1998 Transmitters for industrial process control systems – Part 1: Methods for performance evaluation
  • JB/T 9248-1999 Electromagnetic flowmeters

Terms and Definitions

  • Intrinsic error: Difference between indicated value and true value under reference conditions.
  • Repeatability: Consistency of indications under same conditions.
  • Reference conditions: Standard environmental, power, and fluid conditions for testing.
  • Maximum flow rate: Upper limit for normal operation.
  • Rangeability: Ratio of maximum to minimum flow rate.

Performance Indicators

IndicatorRequirement
Intrinsic error limitClass 0.5: ±0.5%; Class 1.0: ±1.0%; Class 1.5: ±1.5%; Class 2.5: ±2.5%
Repeatability≤ 1/3 of absolute intrinsic error limit
StabilityError change ≤ 1/2 of limit after 7 days continuous operation
Ambient temperature effectOutput change ≤ 1/2 of limit per 10°C change
Power supply effectOutput change ≤ 1/2 of limit for ±10% voltage variation

Test Methods

Static volumetric or master meter methods are used. Test facility uncertainty must be ≤ 1/3 of the EMF's intrinsic error limit. Conditions: 20±5°C, 45%-75% RH, stable power and fluid temperature. At least 5 flow points are tested, including forward/reverse runs.

Inspection Rules

Type inspection covers all performance items every two years or after design changes. Routine inspection includes intrinsic error, repeatability, insulation resistance, and dielectric strength.

Marking, Packaging, and Storage

Product marking includes model, accuracy class, manufacturer, serial number. Packaging protects against moisture and shock. Storage at -40~+70°C, ≤85% RH, no corrosive gases.

Application Recommendations

SIENCUN (Xiechuang Instrument) EMFs comply fully with GB/T 18659-2002. Users should select proper size and lining based on conductivity, temperature, pressure, and flow range. Ensure straight pipe runs (5D upstream, 3D downstream). Regular calibration maintains long-term stability.

Conclusion

GB/T 18659-2002 provides a scientific basis for EMF performance evaluation, essential for measurement quality. Understanding the standard aids correct selection, installation, and maintenance, enhancing process control accuracy and efficiency.